Skip to article frontmatterSkip to article content

Biological effects of radiation

Authors
Affiliations
KU Leuven
KU Leuven

Radiation which interacts with electrons can affect chemical bonds in DNA, and as a result, cause damage to living tissues. The probability that an elementary particle will cause an adverse ionisation is proportional to the energy of the particle deposited in the tissue. In addition, the damage done per unit energy depends on the type of particle.

In diagnostic nuclear medicine, the aim is of course to obtain diagnostic images without or with minimum damage to the patient. Consequently, the amount of activity administered to the patient should be as low as reasonably achievable (ALARA). “Reasonably achievable” means that the radiation emitted by the tracer should still be sufficient to make valuable clinical images.

The amount of radiation deposited in an object is expressed in gray (Gy). A dose of 1 gray is defined as the deposition of 1 joule per kg absorber:

1Gy=1Jkg.1 Gy = 1 \frac{J}{kg}.

To quantify (at least approximately) the amount of damage done to living tissue, the influence of the particle type must be included as well. E.g. it turns out that 1 J deposited by neutrons does more damage than the same energy deposited by photons. To take this effect into account, a quality factor QQ is introduced. Multiplication with the quality factor converts the dose into the dose equivalent. Quality factors are given in Table 1.

Table 1:Quality factor converting dose in dose equivalent (ICRP2007).

Radiation

Q

X-ray, γ-ray,

1

electrons, positrons

1

protons

2

neutrons

2.5 ... 22, depending on energy

α-particals

20

The dose equivalent is expressed in Sv (sievert). Since Q=1Q = 1 for photons, electrons and positrons, we have 1 mSv per mGy in diagnostic nuclear medicine. The natural background radiation is about 2 mSv per year, or about 0.2 μSv per hour.

The older units for radiation dose and dose equivalent are rad and rem:

1 Gy=100 rad1 Sv=100 rem\begin{align} \mbox{1 Gy} &= \mbox{100 rad}\\ \mbox{1 Sv} &= \mbox{100 rem} \end{align}

When the dose to every organ is computed, one can in addition compute an “effective dose”, which is a weighted sum of organ doses. The weights are introduced because damage in one organ is more dangerous than damage in another organ. The most sensitive organs are the gonads (weight about 0.25), the breast, the bone marrow and the lungs (weight about 0.15), the thyroid and the bones (weight about 0.05), see Table 2. The sum of all the weights is 1. The weighted average produces a single value in mSv. The risk of death due to tumor induction is about 5% per “effective Sv” according to report ICRP-60 (International Commission on Radiological Protection (http://www.icrp.org)), but it is of course very dependent upon age (e.g. 14% for children under 10). Research in this field is not finished and the tables and weighting coefficients are adapted every now and then.

Table 2:Organ weight factors to compute the effective dose (ICRP 103, 2007).

organ

weight

total

red marrow, colon, lungs, stomach, breast

0.12

0.60

gonads

0.08

0.08

bladder, liver, esophagus, thyroid

0.04

0.16

brain, skin, salivary glands, bone surfaces

0.01

0.04

remainder (adrenals, extrathoracic region, gallbladder, heart, kidneys, lymphatic nodes, muscle, oral mucosa, pancreas, prostate, small intestine, spleen, thymus, uterus/cervix)

0.00923

0.12

total

1

To obtain the dose equivalent delivered by one organ AA to another organ BB (or to itself, in which case A=BA = B), we have to make the following computation:

DE=iQi  ×  NiA  ×  pi(AB)  ×  Ei  /  MB,\mbox{DE} = \sum_i Q_i \;\times\; N_{iA} \;\times\; p_i(A \rightarrow B) \;\times\; E_i \;/\; M_B,
DE\text{DE}
is the dose equivalent;
ii
denotes a particular emission. Often, multiple photons or particles are emitted during a single radioactive event, and we have to compute the total dose equivalent by summing all contributions;
QiQ_i
is the quality factor for emission ii
NiAN_{iA}
is the total number of particles ii emitted in organ AA
pi(AB)p_i(A \rightarrow B)
is the probability that a particle ii emitted in organ AA will deposit (part of) its energy in organ BB
EiE_i
is the (average) energy that is carried by the particles ii and that can be released in the tissue. For electrons and positrons, this is their kinetic energy. For photons, it is all of their energy.
MBM_B
is the mass of organ BB.

In the following paragraphs, we will discuss these factors in more detail, and illustrate the procedure with two examples.

The particle ii

The ideal tracer for single photon emission would emit exactly one photon in every radioactive event. However, most realistic tracers have a more complex behaviour. Some of them emit several photons at different energies and in different quantities. For example, Iodine-123 has several different ways of decaying (via electron capture) from 53123^{123}_{53}I into 52123^{123}_{52}Te. Each decay scheme produces its own set of photons. Most of these schemes share an energy jump of 159 keV, and the mean number of 159 keV photons emitted per desintegration equals 0.836. In some of the decay schemes, there is an internal conversion electron with an average energy of 127 keV, that comes with a probability of 0.134 per desintegration. In total, 53123^{123}_{53}I uses combinations of 14 different gamma rays, 5 x-rays, 3 internal conversion electrons and 5 Auger electrons to get rid of its excess energy. Each of those has its own probability. In principle, we need to include them all, but it usually suffices to include only the few dominating emissions.

The total number of particles NiAN_{iA}

The number of particles emitted per s changes with time. Due to the finite half life, the radioactivity decays exponentially (see equation (6)). Moreover, the distribution of the tracer molecule in the body is determined by the metabolism, and it changes continuously from the time of injection. Often, the tracer molecule is metabolized: the molecule is cut in pieces or transformed into another one. In those cases, the radioactive atoms may follow one or several metabolic pathways. As a result, it is often very difficult or simply impossible to accurately predict the amount of radioactivity in every organ as a function of time. When a new tracer is introduced, the typical time-activity curves are mostly determined by repeated emission scans in a group of subjects.

Assuming that we know the amount of radioactivity as a function of time, we can compute the total number of particles as

NiA=0niA(t)dt,N_{iA} = \int_0^{\infty} n_{iA}(t) dt,

where niA(t)n_{iA}(t) is the number of particles emitted per s at time tt. For some applications, it is reasonable to assume that the tracer behaviour is dominated by its physical decay. Then we have that

NiA=niA(0)0eln(2)tt1/2dt N_{iA} = n_{iA}(0) \int_0^{\infty} e^{- \ln(2) \frac{t}{t_{1/2}}} dt \nonumber
NiA=niA(0)t1/2ln(2).N_{iA} = n_{iA}(0) \frac{t_{1/2}}{\ln(2)}.

Here niA(0)n_{iA}(0) is the number of particles or photons emitted per s at time 0, and t1/2t_{1/2} is the half life. For a source of 1 MBq at time 0, niA(0)=106n_{iA}(0) = 10^6 per s, since 1 Bq is defined as 1 emission per s.

Often, the tracer is metabolized and sent to the bladder, which implies a decrease of the tracer concentration in most other organs. Therefore, the amount of radioactivity decreases faster than predicted by (6) in these organs. One way to approximate the combined effect of metabolism and physical decay could be to replace the physical half life with a shorter, effective half life in (6). In other cases, it may be necessary to integrate the measured time activity curves numerically.

The probability pi(AB)p_i(A \rightarrow B)

This probability depends on the geometric configuration of the organs and on the attenuation coefficients. The situation is very different for photons on the one hand and electrons and positrons on the other. As illustrated by Table 2 for positrons, the mean path length of an electron and positron in tissue is very short, even for relatively high kinetic energies. Consequently, for these particles, pi(AA)p_i(A \rightarrow A) is close to 1, while pi(AB)p_i(A \rightarrow B) is negligible for different organs AA and BB. The probability that a photon travels from A to B depends on the attenuation along that trajectory. Similarly the probability that it deposits all or a part of its energy in B depends on the attenuation coefficient in organ B.

The energy EiE_i

As mentioned above, EiE_i denotes kinetic energy for electrons and positrons, and the total amount of energy for photons. This energy is usually given in electronvolt. However, to compute the result in Gy, we need the energy in joule. The eV is defined as the amount of energy acquired by an electron if it is accelerated in an electrical field of 1 V. Because joule equals coulomb times volt, we have that

1eV=1.6×1019coulomb×volt=1.6×1019J.1 \mbox{eV} = 1.6 \times 10^{-19} \mbox{coulomb} \times \mbox{volt} = 1.6 \times 10^{-19} \mbox{J}.

Example 1: single photon emission

Consider the configuration shown in Figure 1. Assume that the point source in the left box contains 1 MBq of the isotope 123I. As discussed before, this isotope has a very busy decay scheme, but we will assume that we can ignore all emissions except the following two:

  1. Gamma radiation of 159 keV, with an abundance of 0.84,

  2. Conversion electron of 127 keV, with an abundance of 0.13.

The half life of 123I is 13.0 hours. The boxes are made of plastic or something like that, with an attenuation coefficient of 0.15 cm-1 for photons with energy of 159 keV. The density of the boxes is 1 kg/liter. The boxes are hanging motionless in air, the attenuation of air is negligible. Our task is to estimate the dose that both boxes will receive if we don’t touch this configuration for a few days.

Two objects, one with a radioactive source in the center. The size of the right box is 2 \times 2 \times 4 cm^3.

Figure 1:Two objects, one with a radioactive source in the center. The size of the right box is 2 × 2 × 4 cm3.

We are going to need the mass of the boxes. For the left box we have:

Wleft=43πR3  ×  1kgdm3=0.034  kg,W_{\mbox{left}} = \frac{4}{3} \pi R^3 \; \times \; 1 \frac{\mbox{kg}}{\mbox{dm}^3} = 0.034 \;\mbox{kg},

and for the right box:

Wright=d2L  ×  1kgdm3=0.016  kg.W_{\mbox{right}} = d^2 L\; \times \; 1 \frac{\mbox{kg}}{\mbox{dm}^3} = 0.016 \;\mbox{kg}.

Because a few days lasts several half lifes, a good approximation is to integrate until infinity using equation (6). Consequently, the total number of desintegrations is estimated as

N=1MBq×13 hoursln(2)×3600shour=6.75×1010.N = 1 \mbox{MBq} \times \frac{13\ \mbox{hours}}{\ln(2)} \times 3600 \frac{s}{\mbox{hour}} = 6.75 \times 10^{10}.

This number must be multiplied with the mean number of particles per desintegration:

  • number of photons is 0.84 N = 5.67 × 1010\

  • number of electrons is 0.13 N = 8.78 × 109

The radius of the left box is large compared with the mean path length of the electrons, so the box receives all of their energy: pβ(leftboxleftbox)=1p_{\beta^-}(\mbox{leftbox} \rightarrow \mbox{leftbox}) = 1. Thus, the dose of the left box due to the electrons is:

dose(β,leftbox)=8.78×109  ×  127000 eV  ×1.6×1019 JeV×10.034 kg=5mGy.\mbox{dose}(\beta^-,\mbox{leftbox}) = 8.78 \times 10^9 \; \times \; 127000 \ \mbox{eV} \; \times \frac{1.6 \times 10^{-19} \ \mbox{J}}{\mbox{eV}} \times \frac{1}{0.034 \ \mbox{kg}} = 5 \mbox{mGy}.

For the photons, the situation is slightly more complicated, because not all of them will deposit their energy in the box. Because the box is spherical and the point source is in the center, the amount of attenuating material seen by a photon is the same for every direction. With equation (5) we can compute the probability that a photon will escape without interaction:

pescape=e0Rμdr=eμR=exp(0.15cm  ×  2 cm)=0.74.p_{\mbox{escape}} = e^{- \int_0^R \mu dr} = e^{- \mu R} = \exp(- \frac{0.15}{\mbox{cm}} \; \times \; 2 \ \mbox{cm}) = 0.74.

Thus, there is a chance of 1 - 0.74 = 0.26 that a photon will interact. This interaction could be a Compton event, in which only a part of the photon energy will be absorbed by the box. This complicates the computations considerably. To avoid the problem, we will go for a simple but pessimistic solution: we simply assume that every interaction results in a complete absorption of the photon. That yields for the photon dose to the left box:

dose(γ,leftbox)=5.67×1010  ×159 keV×1.6×1019 JeV×10.034 kg×  0.26=11mGy.\mbox{dose}(\gamma, \mbox{leftbox}) = 5.67 \times 10^{10} \; \times 159\ \mbox{keV} \times \frac{1.6 \times 10^{-19} \ \mbox{J}}{\mbox{eV}} \times \frac{1}{0.034 \ \mbox{kg}} \times \; 0.26 = 11 \mbox{mGy}.

So the total dose to the left box equals about 16 mGy.

For the right box, we can ignore the electrons, they cannot get out of the left box. We already know that a fraction of 0.74 of the photons escapes the left box. Now, we still have to calculate which of those will interact with the right box. We first estimate the fraction traveling along a direction that will bring it into the right box. When observed from the point source position, the right box occupies a solid angle of approximately {math}d^2 / (D + R)^2. The total solid angle equals 4π4 \pi, so the fraction of photons travelling towards the right box equals

pgeometry=d24π(D+R)2=0.0022.p_{\mbox{geometry}} = \frac{d^2}{4 \pi (D + R)^2} = 0.0022.

These photons are travelling approximately horizontally, so most of them are now facing 4 cm of attenuating material. Applying the same pessimistic approach as above yields the fraction that interacts with the material of the right box:

pattenuation=1eμL=0.45.p_{\mbox{attenuation}} = 1 - e^{- \mu L} = 0.45.

Combining all factors yields then number of photons interacting with the right box:

5.67×1010×0.74×0.0022×0.45=4.2×107.5.67 \times 10^{10} \times 0.74 \times 0.0022 \times 0.45 = 4.2 \times 10^7.

Finally, multiplying with the energy and dividing by the mass yields:

dose(γ,rightbox)=4.2×107 ×159 keV×1.6×1019 JeV×10.016 kg=0.067mGy.\mbox{dose}(\gamma, \mbox{rightbox}) = 4.2 \times 10^7\ \times 159\ \mbox{keV} \times \frac{1.6 \times 10^{-19} \ \mbox{J}}{\mbox{eV}} \times \frac{1}{ 0.016 \ \mbox{kg}} = 0.067 \mbox{mGy}.

This is much less than the dose of the left box. The right box is protected by its distance from the source!

Example 2: positron emission tomography

Here, we study the same setup of Figure 1, but now with a point source containing 1 MBq of 18F. This is a positron emitter. From Table 2, we know that the average kinetic energy of the positron is 250 keV. Most of this energy is dissipated in the tissue before the positron annihilates. Thus, there are three contributions to the dose of the left box: the positron and the two photons. The right box can at most be hit by one of the photons.

We assume that in every desintegration, exactly one positron is produced, which annihilates with an electron into two 511 keV. At 511 keV, the attenuation of the boxes is 0.095 cm-1. The half life of 18F is 109 minutes.

The total number of desintegrations is

N=1MBq×109 minln(2)×60smin=9.44×109.N = 1 \mbox{MBq} \times \frac{109\ \mbox{min}}{\ln(2)} \times 60 \frac{s}{\mbox{min}} = 9.44 \times 10^9.

Proceeding in the same way as above, we find the following doses:

dose(β+,leftbox)=9.44×109  ×  250000 eV  ×1.6×1019 JeV×10.034 kg=11mGy.dose(γ,leftbox)=9.44×109  ×  511000 eV  ×1.6×1019 JeV  ×10.034 kg×  0.17=3.9mGy.\begin{align} \mbox{dose}(\beta^+,\mbox{leftbox}) &= 9.44 \times 10^9 \; \times \; 250000 \ \mbox{eV} \; \times \frac{1.6 \times 10^{-19} \ \mbox{J}}{\mbox{eV}} \times \frac{1}{0.034 \ \mbox{kg}} \nonumber\\ &= 11 \mbox{mGy}. \\ \mbox{dose}(\gamma, \mbox{leftbox}) &= 9.44 \times 10^9 \; \times \; 511000 \ \mbox{eV} \; \times \frac{1.6 \times 10^{-19} \ \mbox{J}}{\mbox{eV}} \; \times \frac{1}{0.034\ \mbox{kg}} \times \; 0.17 \nonumber\\ &= 3.9 \mbox{mGy}. \end{align}

The total dose then equals dose(β+,leftbox)+2  dose(γ,leftbox)\mbox{dose}(\beta^+,\mbox{leftbox}) + 2 \; \mbox{dose}(\gamma, \mbox{leftbox}) = 18.8 mGy.

For the right box, we first estimate the number of interacting photons (remember that there are two photons per desintegration):

2N pescape pgeometry pattenuation=2N×0.83×0.0022×0.32=1.09×107.2N \ p_{\mbox{escape}}\ p_{\mbox{geometry}}\ p_{\mbox{attenuation}} = 2N \times 0.83 \times 0.0022 \times 0.32 = 1.09 \times 10^7.

And for the dose to the right box we get:

dose(γ,rightbox)=1.09×107  ×  511000 eV  ×1.6×1019 JeV×10.016 kg=0.056mGy.\mbox{dose}(\gamma, \mbox{rightbox}) = 1.09 \times 10^7\; \times \; 511000 \ \mbox{eV} \; \times \frac{1.6 \times 10^{-19} \ \mbox{J}}{\mbox{eV}} \times \frac{1}{0.016\ \mbox{kg}} = 0.056 \mbox{mGy}.

Internal dosimetry calculations

Internal dosimetry calculations use the same principle as in the examples above to compute the dosage to every organ. Because there are many organs, and because the activity in any organ contributes to the dose in any other organ, the computations are complex and tedious. This is one of the reasons why dedicated software has been developed. With the fast computers we have today, the calculations can be carried out more accurately. E.g. it is possible to take the effects of Compton scatter into account using Monte Carlo techniques.

Of course, we still have to provide a meaningful input to the software. In principle, we should input the precise anatomy of the patient. Because this is usually not available, this problem is avoided by using a few standard geometries, which model the human anatomy, focussing on the most senstive organs. It follows that the resulting doses should only be considered as estimates, not as accurate numbers.

We also have to provide the tracer concentration as a function of time for all organs. For known tracers, the software may contain typical residence times; for new ones, measurements in humans have to be carried out to estimate the dosimetry.